User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Power Mate ES-28H:Power Supply DATE: 17-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 9 NUMBER OF LINES: 94 CONFIGURATION: Datron 1281 STANDARD: DTB Metered Variac STANDARD: Transistor Devices DLP 50 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- P F 1.003 HEAD INITIAL CONDITIONS 1.004 DISP Set DTB Metered Variac to the OFF position. 1.005 DISP Connect the DTB Metered Variac to the UUT as follows: 1.005 DISP AC output to UUT AC terminal connections. 1.006 DISP Connect the UUT and the DLP 50 as follows: 1.006 DISP ES.28H TO DLP 50 1.006 DISP OUTPUT +V ............> INPUT (+) 1.006 DISP OUTPUT -V ............> INPUT (-) 1.007 DISP Connect the UUT and the 1281 as follows: 1.007 DISP ES.28H TO 1281 1.007 DISP OUTPUT (+) ............> INPUT HIGH 1.007 DISP OUTPUT (-) ............> INPUT LO 1.008 DISP Set the variac to 115 VAC. 1.009 STD DTB Metered Variac 1.010 HEAD {* VOLTAGE ACCURACY TEST *} 1.011 JMP 2.001 1.012 EVAL 2.001 IEEE [@1281] DCV AUTO 2.002 IEEE [@1281] [D3000]RDG?[I] 2.003 MATH MEM1 = 28.00 2.004 MEME 2.005 MEMC V 10% 0.00U 3.001 HEAD {* LOAD REGULATION TEST @ 10 AMPS *} 3.002 JMP 4.001 3.003 EVAL 4.001 DISP Set the DLP 50 front panel controls as follows: 4.001 DISP VOLTS RANGE ........................................ 60V 4.001 DISP AMPS RANGE ......................................... 18A 4.001 DISP MODE ............................................. 0.30A 4.001 DISP DC switch .......................................... OFF 4.001 DISP LOAD ADJUST COARSE/FINE ...................... fully ccw 4.001 DISP AC ON switch ........................................ ON 4.002 STD Transistor Devices DLP 50 4.003 DISP Set the DLP 50 front panel controls as follows: 4.003 DISP DC switch ........................................... ON 4.004 DISP Adjust DLP 50 COARSE and FINE LOAD ADJUST controls 4.004 DISP for a DLP 50 reading of 10 Amps. 4.005 IEEE [@1281] DCV AUTO 4.006 IEEE [@1281] [D3000]RDG?[I] 4.007 MEME 4.008 DISP Set the DLP 50 front panel controls as follows: 4.008 DISP DC switch .......................................... OFF 4.009 IEEE [@1281] DCV AUTO 4.010 IEEE [@1281] [D3000]RDG?[I] 4.011 MATH MEM = (MEM - MEM1) 4.012 MATH MEM1 = 0.00 4.013 MEME 4.014 MEMC mV 57.00U FULLtoNO 5.001 HEAD {* LINE REGULATION TEST @ 10 AMPS *} 5.002 JMP 6.001 5.003 EVAL 6.001 DISP Set the DLP 50 front panel controls as follows: 6.001 DISP DC switch ........................................... ON 6.002 DISP Set the variac to 105 VAC. 6.003 IEEE [@1281] DCV AUTO 6.004 IEEE [@1281] [D3000]RDG?[I] 6.005 MEME 6.006 DISP Set the variac to 125 VAC. 6.007 IEEE [@1281] DCV AUTO 6.008 IEEE [@1281] [D3000]RDG?[I] 6.009 MATH MEM = (MEM - MEM1) * 1000 6.010 MATH MEM1 = 0.00 6.011 MEME 6.012 MEMC mV 57.00U 105to125VAC 7.001 DISP Set the variac to 115 VAC. 7.002 HEAD {* RIPPLE AND NOISE TEST *} 7.003 JMP 8.001 7.004 EVAL 8.001 IEEE [@1281]ACV AUTO 8.002 IEEE [@1281][D5000]RDG?[I] 8.003 MATH MEM=(MEM*2.82)*1000 8.004 MATH MEM1 = 0.00 8.005 MEME 8.006 MEMC mVP.P +50.00U 9.001 DISP Set the DLP 50 front panel controls as follows: 9.001 DISP DC switch .......................................... OFF 9.002 DISP Set DTB Metered Variac to the OFF position. 9.003 END